NGC3 Ion Gauge process controller with Active Gauge input. Supports fully automatic gauge operation to continuously display pressure across the full range of 1200 mBar to 3 x 10-11 mBar.
| Order code | Description | Price excl. VAT and transportation cost |
|---|---|---|
| ngc3 | Ion Gauge Controller | £1,795.00 |
AML AIG1xG are recommended. Bayard-Alpert gauges from many other manufacturers are suitable without adjustment other than sensitivity.
From 1 x 10-3 to below 3 x 10-11 mBar with a UHV gaugehead with tungsten filaments. The lower limit is dependent on gaugehead, cable construction, cable length and conditions of use. The upper limit is determined by the acceptable life of the filament and my be extended by the use of thoria or yttria-coated iridium filaments.
Determined principally by the gauge-head: controller errors are much smaller. Emission at 0.5 mA is recommended. Electrometer logarithmic conformance at 21°C, <1% at 1 mA to 10 pA, <10% at 10 pA to 2 pA.
Grid: +200 V in emission, +500 V at ≤60 mA in degas.
Filament: +50 V bias, ≤12 V at ≤4.2 A (Tungsten), ≤2.6 A (Iridium) with filament power limited to 30 W maximum.
AML types PVU and PVB.
A constant-voltage bridge circuit reduces contamination at high pressures. AML Pirani gaugeheads may be exchanged or extension leads may be connected without adjustments being necessary.
Self-powered or Active Gauge with +10 V full-scale output. Format selectable between linear (1, 10, 100, 1000 mBar or Torr full scale) or log (1 V/decade, 0.5 V at 1 x 10-6 mBar). The instrument provides a regulated +24 Vdc supply, 1 W maximum, protected by a 50 mA self-resetting fuse to power a connected Active gauge.
4 x single-pole, change-over. 5 A at 240 V maximum.
Independently assignable to any gauge.
Mineral-insulated K-type with miniature flat-pin connector.
Settable bake temperature (50°C to 250°C), bake time (1 hour to 90 hours) and overpressure limit.
RS232C
1200, 2400, 4800 or 9600 (Default) baud, 8 data bits, 1 stop bit, no parity, no handshaking.
Scientific notation (1 or 2 decimal place resolution) or bar-graph displays in mBar, Torr or Pascal.
Whole values in pA, nA, μA and mA.
5°C to 35°C for specified performance. Incoming air temperature is measured and displayed. Operation is inhibited at >40°C.
100 V to 240 V nominal at 48 to 65 Hz, without adjustment.
<20 W idling, <75 W in emission.
Width: 19″ full-width rack (482.6 mm), Height: 1U (44.45 mm), Depth 270 mm.
2.7 kg